PLAN+ TEST OBJECT

A dual phantom to test the precision of 4D and 6D radiotherapy treatment tables and the alignment of the lasers positioning system on a CT-scanner

Quote

 

OTP-PLAN+ is a dual phantom:

 

  • Using 2D-MV portal images of the back part of the phantom, you evaluate the precision of the 3 translations and 3 rotations of the 4D or 6D couches used in radiation therapy (please see TABLE module).
 
  • The front part of the phantom integrates all the characteristics of the OTP-PLAN phantom and therefore allows you to control the alignment of the laser positioning system on a scanner used in radiology or radiation therapy (please see PLAN module).

 

Specifications

Less than 30 minutes is all the time you need to perform the entire controlinstallation of the test object, image acquisition for 15 rotations and translations, analysis, editing of 6 different PDF analysis reports and a trend curve for each tested parameter.

 

The TABLE module enables you to optimally fulfil requirements of the French National Agency for the Safety of Medicines (ANSM) decision dated 28/02/23 regulating the procedures for quality control of external radiation therapy and radiosurgery devices: point 5.1, 5.2 and 5.3 of the annex.